5.2-4 NULL setting (Protective conductor resistance range)
According to below figure, set the range switch to the range for protective
conductor resistance (either 20Ω+ or 20Ω- is OK). Press NULL button after
short-circuited PE/2mA and RPE -Test lead terminals with RPE-Test Lead.
Make sure that the " " mark appears on the LCD and indicated value is
"0.00Ω" ; Data is corrected.
CAUTION
● When pressing NULL button, the indicated value is stored as a reference
value. After that, the difference between the reference value (stored
value) and measuring value is indicated on the LCD.
● The " " mark is displayed on the LCD while the NULL function is
operating. Each time pressing NULL button, NULL function shall be
turned on and off.
● NULL function can not be released by changing the range while it is
operating. To release the function, power off the instrument.
● NULL function works on both 20Ω+ range and 20Ω- range, therefore,
there is no need to do NULL setting at both ranges.
● For the measuring method shown in the following clause 6.3 and 6.4,
NULL setting is required also for an optional cord.