5.2-4 NULL setting (Protective conductor resistance range)
According to below figure, set the range switch to the range for protective
conductor resistance (either 20Ω+ or 20Ω- is OK). Press NULL button after
short-circuited PE/2mA and RPE -Test lead terminals with RPE-Test Lead.
Make sure that the " " mark appears on the LCD and indicated value is
"0.00Ω" ; Data is corrected.
CAUTION
When pressing NULL button, the indicated value is stored as a reference
value. After that, the difference between the reference value (stored
value) and measuring value is indicated on the LCD.
The " " mark is displayed on the LCD while the NULL function is
operating. Each time pressing NULL button, NULL function shall be
turned on and off.
NULL function can not be released by changing the range while it is
operating. To release the function, power off the instrument.
NULL function works on both 20Ω+ range and 20Ω- range, therefore,
there is no need to do NULL setting at both ranges.
For the measuring method shown in the following clause 6.3 and 6.4,
NULL setting is required also for an optional cord.